07-790R-014NOTES "C" (CLINAC)
1. CONTRACTOR SHALL FURNISH ALL UNIT MANUFACTURERS ELECTRICAL PROVISIONS
REQUIRED FOR A COMPLETE AND SATISFACTORY INSTALLATION AND FINAL CONNECTION
OF THIS MEDICAL EQUIPMENT. VERIFY ALL UNIT MANUFACTURER REQUIREMENTS
BEFORE ROUGHING -IN.
2. ALL WORK ON MEDICAL EQUIPMENT AREAS SHALL COMPLY WITH THE LATEST EDITION
OF THE NATIONAL ELECTRICAL CODE AND APPLICABLE STATE, CITY AND LOCAL CODES
AND ORDINANCES.
3. ALL CONDUCTORS SERVING MEDICAL EQUIPMENT OR MEDICAL SYSTEMS SHALL BE
COPPER AND THHN /THWN TYPE EXCEPT OTHERWISE NOTED. RACEWAYS SHALL BE
ELECTRIC METALLIC TUBING (EMT) (WHEN APPLICABLE). WHEN HARING IS PROVIDED
FOR UNIT MANUFACTURER INSTALLER FINAL CONNECTION, LEAVE A MINIMUM OF 5FT
(OR 1OFT AS REQUIRED) WIRE TAILS AT OUTLET POINT, WITH WIRE IDENTIFICATION AT
BOTH ENDS.
4. CONTRACTOR SHALL VERIFY AT FIELD ALL WIREWAY /RACEWAY
ROUTING - INSTALLATION, TO COMPLY WITH MAXIMUM PERMISSIBLE LENGTHS FOR
PRECUT OR STANDARD LENGTHS OF CONDUCTORS FURNISHED BY OTHERS, SEE
MAXIMUM LENGTHS WITH UNIT MANUFACTURER SHOP DRAWINGS CONTROL DIAGRAMS
5. FOR TELEPHONE AND DATA REQUIREMENTS REFER TO POWER FLOOR PLAN DRAWING
E -2 AND UNIT MANUFACTURER SHOP DRAWINGS.
6. PROVIDE A DEDICATED EQUIPMENT GROUND WIRE (GREEN) IN ALL BRANCH WRING
SERVING MEDICAL EQUIPMENT, TO COMPLY WITH N.E.C. 250.122 REQUIREMENTS FOR
ELECTRICAL BOXES, RACEWAYS AND NON - MEDICAL EQUIPMENT. (G) = GROUND WIRE
7. ALL CONDUIT BEND TO 6 TIMES THE DIAMETER AND DO NOT EXCEED 270 DEG
BENDS PER CONDUIT RUN.
8. CONTRACTOR SHALL NEATLY LABEL ALL OUTLETS WITH PANEL AND CIRCUIT
SERVED. ALL PANELS, JUNCTION BOXES, PULL BOXES, DISCONNECTS AND
CONDUITS SHALL BE LABELED ACCORDINGLY.
B -24
SEC PANEL
B -22
TEL SYS.
CT TRASTECTOR STE
XFMR -UNE CONDIT],
208V, 3PH, 3W INPU
120/208V, 3PH, OUT
277/48OV, 3PH, OU1
PANEL "DPA"
120/208V, 3PH, 4
75` �VA XFMR
(STEP UP)
(ABO�j_
TELEPHO r
ELECTRICAL RM.
3
PANEL "C'
A
120/208V, 3PH, 4W
(MEO EQUIP.)
CORRIDOR
RETRACTA13LE
VINYL CURTAIN
14 SEE IBASE FRAME CABLE ACCESS
DETAILS VARIAN SILHOUETTE EDITTO
CLINAC INSTALLATION MANUAL REV
PG 7.35.0 AND REV 3 PG 7.36.0
*HD {
( C -17
I `
C -1
1 1
J I .
TREATMENT
fOR l OUNTTNG HT AND EXACT LOCATION
' SEE IN -ROOM MONITOR INSTALLATION i
DETAILS VARIAN SILHOUETTE EDITION
CLINAC MANUAL REV 2 PG 7.33.0
4 ! *HD
18 8 C -158
GA
io *HID C -1
-15 C - -1 3 8
*HD
IC
' ,x,�i� x ��,
c -/ l T X
9 C -7 C -9 C -13 C -11 C -13 C -11 C -19
D *HD *HO *HO *HD *HD *HD *HD
8 1 5 10
CLINAC OBI TRILOG MLC TYP.
ONTROL EQUIPMENT REQUIREMENTS AND
IS EXACT; IL.g N8 EF R TO UNIT
ACTURE&.L7P1 AL(E�T VARIAN
ETTE EDITION CLINAC MANUAL REV 2 PG 7.27.0
CONDUIT ROUTING SHOWN "FOR REFERENCE ONLY'.
ELECTRICAL CONTRACTOR SHALL STRICTLY COORDINATE
ALL ELECTRICAL RACEWAYS ROUTING, "IN- AND -OUT" AND/OR
SERVING THE VAULT / CLINAC AREAS, NVITH SHOP DRAWINGS FOR
VAULT FABRICATION BY ATOMIC INTERNATIONAL INC. AND
UNIT MANUFACTURER INSTALLER. COORDINATE ALL EQUIPMENT
PROVISIONS EXACT FINAL LOCATIONS AND MOUNTING HEIGHTS
WITH MEDICAL EQUIPMENT AT THE SITE BEFORE ROUGING -IN.
(�& FLOOR PLAN _ CLINAC VAULT
SCALE: 1/2" = 1' -0"
*HD
C -11
C -11
*HD
C-11
*HD
F- ( YP.)
8 LOW
eC UNTER
I
( )
SYMBOL
V
DEDICATED DIRECT TELEPHONE IN
THE LOCATION OF THE VARIAN SY
O
OUTLET, REFER TO TELEPHONE AN
Fes'
Y
/ /
BEAM -ON WARNING LIGHT. PROVIDI
O
REQUIRED TO BLWK WHEN X -RAY
U
(24190, 250MA), INSIDE VAULT "I
POINT IN THE CLINAC ROOM. PRO
J
RNAITE CONDDUIT ACCESS
W
SILHOUETTE CABINET.
O
14 SEE IBASE FRAME CABLE ACCESS
DETAILS VARIAN SILHOUETTE EDITTO
CLINAC INSTALLATION MANUAL REV
PG 7.35.0 AND REV 3 PG 7.36.0
*HD {
( C -17
I `
C -1
1 1
J I .
TREATMENT
fOR l OUNTTNG HT AND EXACT LOCATION
' SEE IN -ROOM MONITOR INSTALLATION i
DETAILS VARIAN SILHOUETTE EDITION
CLINAC MANUAL REV 2 PG 7.33.0
4 ! *HD
18 8 C -158
GA
io *HID C -1
-15 C - -1 3 8
*HD
IC
' ,x,�i� x ��,
c -/ l T X
9 C -7 C -9 C -13 C -11 C -13 C -11 C -19
D *HD *HO *HO *HD *HD *HD *HD
8 1 5 10
CLINAC OBI TRILOG MLC TYP.
ONTROL EQUIPMENT REQUIREMENTS AND
IS EXACT; IL.g N8 EF R TO UNIT
ACTURE&.L7P1 AL(E�T VARIAN
ETTE EDITION CLINAC MANUAL REV 2 PG 7.27.0
CONDUIT ROUTING SHOWN "FOR REFERENCE ONLY'.
ELECTRICAL CONTRACTOR SHALL STRICTLY COORDINATE
ALL ELECTRICAL RACEWAYS ROUTING, "IN- AND -OUT" AND/OR
SERVING THE VAULT / CLINAC AREAS, NVITH SHOP DRAWINGS FOR
VAULT FABRICATION BY ATOMIC INTERNATIONAL INC. AND
UNIT MANUFACTURER INSTALLER. COORDINATE ALL EQUIPMENT
PROVISIONS EXACT FINAL LOCATIONS AND MOUNTING HEIGHTS
WITH MEDICAL EQUIPMENT AT THE SITE BEFORE ROUGING -IN.
(�& FLOOR PLAN _ CLINAC VAULT
SCALE: 1/2" = 1' -0"
*HD
C -11
C -11
*HD
C-11
*HD
F- ( YP.)
8 LOW
eC UNTER
I
( )
NOTES:
1. CONTRACTOR SHALL PROVIDE ALL REQUIRED PROVISIONS FOR THE COMPLETE INSTALLATION OF THE WARNING LIGHTS SYSTEM, LASER POSITIONING LIGHTS AND RELAY, CCTV SYSTEM, INTERCOM,
DETECTION SYSTEM, UNDER VOLTAGE RELAY, DOOR INTERLOCKS, EM SHUT -OFF SW, RM LTS RELAY, AND CLWAC PERIPHERALS AS SIHOWN. CONTRACTOR SHALL PULL VARIAN SUPPLIED CABLES
WHEN APPLICABLE, REFER TO INTERCONNECTION WRING DIAGRAM FOR SILHOUETTE EDITION CLINAC REV 3, PG 7.38.0 AND SHOWN ON CUSTOMER SUPPLIED WE COORDINATE WITH MEDICAL
EQUIPMENT INSTALLER AT FIELD BEFORE ROUGHING -IN.
2. ALL RACEWAYS FILLING AND PULL AND JUNCTION BOXES SIZES SHALL COMPLY WITH N.E.O. ART. 314, VERIFY ALL SIZES WITH UNIT MANUFACTURER
SHOP DRAWINGS AND FIELD REQUIREMENTS AND CONDITIONS BEFORE ROUGHING -IN.
NOTES:
1. VERIFY ALL SIZES WITH UNIT MANUFACTURER SHOP DRAWING AND FIELD REQUIREMENTS AND CONDITIONS BEFORE ROUGHING -IN. REFER TO TO UNIT
MANUFACTURER'S CONTROL DIAGRAMS.
2. ALL CONDUIT MAXIMUM LENGTHS SHALL COMPLY WITH MANUFACTURER REQUIREMENTS AS SHOWN ON UNIT MANUF DRAWNG% COORDINATE
AT FIELD THE BEST ROUTING TO COMPLY.
3. RACEWAYS FILLING AND PULL AND JUNCTION BOXES SIZES SHALL COMPLY WTH N.E.C. ART. 314,
C;1TfUIILOING Mt
DEPT
O
ELE,CTHICAL SECTION
A PPROVED 6y: L6 '
r,AT E
SYMBOL
V
DEDICATED DIRECT TELEPHONE IN
THE LOCATION OF THE VARIAN SY
O
OUTLET, REFER TO TELEPHONE AN
Fes'
VARIAN REQUIREMENTS REV 2 PG
BEAM -ON WARNING LIGHT. PROVIDI
O
REQUIRED TO BLWK WHEN X -RAY
U
(24190, 250MA), INSIDE VAULT "I
POINT IN THE CLINAC ROOM. PRO
O
WARNING LIGHTS RELAY A 1Y x 1.
W
CABINET. PROVIDE A POTTER -BRL
O
PROVIDE CCTV CAMERA SYSTEM 2
RECEPTACLES AS SHOWN. SYSTEM
U.P.S. UNIT TO CAMERA AND TO C
0
PROVIDE TWO -WAY PATIENT MOAT
O
WITH PUSH -TO TALK CAPABILITIES.
DOOR OPENER, 12OV, 1
OMOTORIZED
g
ELECTRICAL CONTRACTOR.. PROVIL
3�Q
LASER POSITIONING LIGHT, POW-RE
O
CORD AND CONNECTED TO A OUR
e
RELAY J-BOX, PER UNIT MANUFAC
LL.
= M
IN -ROOM MONITOR , PROVIDE A SI
O
VERIFY SIZE REQUIREMENTS AT FIE
W
FRIO PER DIRECTIONS MEDICAL EC
O
CONTROL EQUIPMENT PULL BOX 18
FROM PHYSICAL DAMAGE AND LOC
COMBO TELEPHONE/DATA SIGNAL
10
VOICE AND DATA DWG E -4. SEE
ALL SERVERS AND WORKSTATION E
11
COMBO TEIEPHONE/DATA SIGNAL 1
VOICE AND DATA DWG E -4.
ON -BOARD IMAGER MAIN CIRCUIT
12
CONSOLE CONTRACTOR SHALL U
INSTALLED IN A LOCKOUT /TAGOUT
MANUfACTUREO BY G.E. CATJ OBI+
MAIN CIRCUIT BREAKER, IN NEMA '
f3
SHALL LABEL DISCONNECT AS 'MA
150A UNDER VOLTAGE RELEASE BF
MANUFACTURER INSTALLATION GUN
PROVIDE A 1Yx2Vx10' FLOOR RE,
14
RECOMMENDATIONS SEE DETAIL RE
PROVIDE PROTECTION TO PREVENT
EXTERNAL EMERGENCY SHUT -OFF
15
MANUAL RESET) IN ADDITION TO T
WIRING CONTROL DIAGRAMS.
ROOM SINGLE MASTER SWITCH WAIT
ig
TELEVISION CAN HAVE AN OPERAT
CAN BE SUBORDINATED TO MIX
OPERATION, REFER TO MANUFACTL
17
SAFETY DOOR INTERLOCK SWITCH.
NORMALLY OPEN TYPE SWTCHES.
t8
INDEPENDENT RADIATION DETECTOF
1.50 J -BOXES AT DETECTOR AND
LOCATIONS BEFORE ROUGHING -IN
NOTES:
1. CONTRACTOR SHALL PROVIDE ALL REQUIRED PROVISIONS FOR THE COMPLETE INSTALLATION OF THE WARNING LIGHTS SYSTEM, LASER POSITIONING LIGHTS AND RELAY, CCTV SYSTEM, INTERCOM,
DETECTION SYSTEM, UNDER VOLTAGE RELAY, DOOR INTERLOCKS, EM SHUT -OFF SW, RM LTS RELAY, AND CLWAC PERIPHERALS AS SIHOWN. CONTRACTOR SHALL PULL VARIAN SUPPLIED CABLES
WHEN APPLICABLE, REFER TO INTERCONNECTION WRING DIAGRAM FOR SILHOUETTE EDITION CLINAC REV 3, PG 7.38.0 AND SHOWN ON CUSTOMER SUPPLIED WE COORDINATE WITH MEDICAL
EQUIPMENT INSTALLER AT FIELD BEFORE ROUGHING -IN.
2. ALL RACEWAYS FILLING AND PULL AND JUNCTION BOXES SIZES SHALL COMPLY WITH N.E.O. ART. 314, VERIFY ALL SIZES WITH UNIT MANUFACTURER
SHOP DRAWINGS AND FIELD REQUIREMENTS AND CONDITIONS BEFORE ROUGHING -IN.
NOTES:
1. VERIFY ALL SIZES WITH UNIT MANUFACTURER SHOP DRAWING AND FIELD REQUIREMENTS AND CONDITIONS BEFORE ROUGHING -IN. REFER TO TO UNIT
MANUFACTURER'S CONTROL DIAGRAMS.
2. ALL CONDUIT MAXIMUM LENGTHS SHALL COMPLY WITH MANUFACTURER REQUIREMENTS AS SHOWN ON UNIT MANUF DRAWNG% COORDINATE
AT FIELD THE BEST ROUTING TO COMPLY.
3. RACEWAYS FILLING AND PULL AND JUNCTION BOXES SIZES SHALL COMPLY WTH N.E.C. ART. 314,
C;1TfUIILOING Mt
DEPT
O
ELE,CTHICAL SECTION
A PPROVED 6y: L6 '
r,AT E
ul3
3�
a
U
O`
l
c
LL..
V
W
Fes'
S
U
W
(�
N
O
W
w
>
H
0
P4
U
N N
Z
`/
U
3�Q
w0co
e
LL.
= M
Z
U
C4
W
v
Z�~�
W
0
CO
VJ
ul3
3�
a
U
O`
l
c
LL..
0
VJ
Z
W
(�
N
O
W
w
>
H
0
P4
`/
U
w w
w0co
e
LL.
= M
Z
U
Q
Z�~�
W
Q
CO
VJ
00
0-9
0
12l
DATE: 03/14/07
PROD NO: -
DWG BY: JEB
REV: -
04/06/07
A 212/10/2007
IT- REVISION
02/04/2008
3 CT -CHG.
E -3
a