07-790R-014NOTES "C" (CLINAC) 1. CONTRACTOR SHALL FURNISH ALL UNIT MANUFACTURERS ELECTRICAL PROVISIONS REQUIRED FOR A COMPLETE AND SATISFACTORY INSTALLATION AND FINAL CONNECTION OF THIS MEDICAL EQUIPMENT. VERIFY ALL UNIT MANUFACTURER REQUIREMENTS BEFORE ROUGHING -IN. 2. ALL WORK ON MEDICAL EQUIPMENT AREAS SHALL COMPLY WITH THE LATEST EDITION OF THE NATIONAL ELECTRICAL CODE AND APPLICABLE STATE, CITY AND LOCAL CODES AND ORDINANCES. 3. ALL CONDUCTORS SERVING MEDICAL EQUIPMENT OR MEDICAL SYSTEMS SHALL BE COPPER AND THHN /THWN TYPE EXCEPT OTHERWISE NOTED. RACEWAYS SHALL BE ELECTRIC METALLIC TUBING (EMT) (WHEN APPLICABLE). WHEN HARING IS PROVIDED FOR UNIT MANUFACTURER INSTALLER FINAL CONNECTION, LEAVE A MINIMUM OF 5FT (OR 1OFT AS REQUIRED) WIRE TAILS AT OUTLET POINT, WITH WIRE IDENTIFICATION AT BOTH ENDS. 4. CONTRACTOR SHALL VERIFY AT FIELD ALL WIREWAY /RACEWAY ROUTING - INSTALLATION, TO COMPLY WITH MAXIMUM PERMISSIBLE LENGTHS FOR PRECUT OR STANDARD LENGTHS OF CONDUCTORS FURNISHED BY OTHERS, SEE MAXIMUM LENGTHS WITH UNIT MANUFACTURER SHOP DRAWINGS CONTROL DIAGRAMS 5. FOR TELEPHONE AND DATA REQUIREMENTS REFER TO POWER FLOOR PLAN DRAWING E -2 AND UNIT MANUFACTURER SHOP DRAWINGS. 6. PROVIDE A DEDICATED EQUIPMENT GROUND WIRE (GREEN) IN ALL BRANCH WRING SERVING MEDICAL EQUIPMENT, TO COMPLY WITH N.E.C. 250.122 REQUIREMENTS FOR ELECTRICAL BOXES, RACEWAYS AND NON - MEDICAL EQUIPMENT. (G) = GROUND WIRE 7. ALL CONDUIT BEND TO 6 TIMES THE DIAMETER AND DO NOT EXCEED 270 DEG BENDS PER CONDUIT RUN. 8. CONTRACTOR SHALL NEATLY LABEL ALL OUTLETS WITH PANEL AND CIRCUIT SERVED. ALL PANELS, JUNCTION BOXES, PULL BOXES, DISCONNECTS AND CONDUITS SHALL BE LABELED ACCORDINGLY. B -24 SEC PANEL B -22 TEL SYS. CT TRASTECTOR STE XFMR -UNE CONDIT], 208V, 3PH, 3W INPU 120/208V, 3PH, OUT 277/48OV, 3PH, OU1 PANEL "DPA" 120/208V, 3PH, 4 75` �VA XFMR (STEP UP) (ABO�j_ TELEPHO r ELECTRICAL RM. 3 PANEL "C' A 120/208V, 3PH, 4W (MEO EQUIP.) CORRIDOR RETRACTA13LE VINYL CURTAIN 14 SEE IBASE FRAME CABLE ACCESS DETAILS VARIAN SILHOUETTE EDITTO CLINAC INSTALLATION MANUAL REV PG 7.35.0 AND REV 3 PG 7.36.0 *HD { ( C -17 I ` C -1 1 1 J I . TREATMENT fOR l OUNTTNG HT AND EXACT LOCATION ' SEE IN -ROOM MONITOR INSTALLATION i DETAILS VARIAN SILHOUETTE EDITION CLINAC MANUAL REV 2 PG 7.33.0 4 ! *HD 18 8 C -158 GA io *HID C -1 -15 C - -1 3 8 *HD IC ' ,x,�i� x ��, c -/ l T X 9 C -7 C -9 C -13 C -11 C -13 C -11 C -19 D *HD *HO *HO *HD *HD *HD *HD 8 1 5 10 CLINAC OBI TRILOG MLC TYP. ONTROL EQUIPMENT REQUIREMENTS AND IS EXACT; IL.g N8 EF R TO UNIT ACTURE&.L7P1 AL(E�T VARIAN ETTE EDITION CLINAC MANUAL REV 2 PG 7.27.0 CONDUIT ROUTING SHOWN "FOR REFERENCE ONLY'. ELECTRICAL CONTRACTOR SHALL STRICTLY COORDINATE ALL ELECTRICAL RACEWAYS ROUTING, "IN- AND -OUT" AND/OR SERVING THE VAULT / CLINAC AREAS, NVITH SHOP DRAWINGS FOR VAULT FABRICATION BY ATOMIC INTERNATIONAL INC. AND UNIT MANUFACTURER INSTALLER. COORDINATE ALL EQUIPMENT PROVISIONS EXACT FINAL LOCATIONS AND MOUNTING HEIGHTS WITH MEDICAL EQUIPMENT AT THE SITE BEFORE ROUGING -IN. (�& FLOOR PLAN _ CLINAC VAULT SCALE: 1/2" = 1' -0" *HD C -11 C -11 *HD C-11 *HD F- ( YP.) 8 LOW eC UNTER I ( ) SYMBOL V DEDICATED DIRECT TELEPHONE IN THE LOCATION OF THE VARIAN SY O OUTLET, REFER TO TELEPHONE AN Fes' Y / / BEAM -ON WARNING LIGHT. PROVIDI O REQUIRED TO BLWK WHEN X -RAY U (24190, 250MA), INSIDE VAULT "I POINT IN THE CLINAC ROOM. PRO J RNAITE CONDDUIT ACCESS W SILHOUETTE CABINET. O 14 SEE IBASE FRAME CABLE ACCESS DETAILS VARIAN SILHOUETTE EDITTO CLINAC INSTALLATION MANUAL REV PG 7.35.0 AND REV 3 PG 7.36.0 *HD { ( C -17 I ` C -1 1 1 J I . TREATMENT fOR l OUNTTNG HT AND EXACT LOCATION ' SEE IN -ROOM MONITOR INSTALLATION i DETAILS VARIAN SILHOUETTE EDITION CLINAC MANUAL REV 2 PG 7.33.0 4 ! *HD 18 8 C -158 GA io *HID C -1 -15 C - -1 3 8 *HD IC ' ,x,�i� x ��, c -/ l T X 9 C -7 C -9 C -13 C -11 C -13 C -11 C -19 D *HD *HO *HO *HD *HD *HD *HD 8 1 5 10 CLINAC OBI TRILOG MLC TYP. ONTROL EQUIPMENT REQUIREMENTS AND IS EXACT; IL.g N8 EF R TO UNIT ACTURE&.L7P1 AL(E�T VARIAN ETTE EDITION CLINAC MANUAL REV 2 PG 7.27.0 CONDUIT ROUTING SHOWN "FOR REFERENCE ONLY'. ELECTRICAL CONTRACTOR SHALL STRICTLY COORDINATE ALL ELECTRICAL RACEWAYS ROUTING, "IN- AND -OUT" AND/OR SERVING THE VAULT / CLINAC AREAS, NVITH SHOP DRAWINGS FOR VAULT FABRICATION BY ATOMIC INTERNATIONAL INC. AND UNIT MANUFACTURER INSTALLER. COORDINATE ALL EQUIPMENT PROVISIONS EXACT FINAL LOCATIONS AND MOUNTING HEIGHTS WITH MEDICAL EQUIPMENT AT THE SITE BEFORE ROUGING -IN. (�& FLOOR PLAN _ CLINAC VAULT SCALE: 1/2" = 1' -0" *HD C -11 C -11 *HD C-11 *HD F- ( YP.) 8 LOW eC UNTER I ( ) NOTES: 1. CONTRACTOR SHALL PROVIDE ALL REQUIRED PROVISIONS FOR THE COMPLETE INSTALLATION OF THE WARNING LIGHTS SYSTEM, LASER POSITIONING LIGHTS AND RELAY, CCTV SYSTEM, INTERCOM, DETECTION SYSTEM, UNDER VOLTAGE RELAY, DOOR INTERLOCKS, EM SHUT -OFF SW, RM LTS RELAY, AND CLWAC PERIPHERALS AS SIHOWN. CONTRACTOR SHALL PULL VARIAN SUPPLIED CABLES WHEN APPLICABLE, REFER TO INTERCONNECTION WRING DIAGRAM FOR SILHOUETTE EDITION CLINAC REV 3, PG 7.38.0 AND SHOWN ON CUSTOMER SUPPLIED WE COORDINATE WITH MEDICAL EQUIPMENT INSTALLER AT FIELD BEFORE ROUGHING -IN. 2. ALL RACEWAYS FILLING AND PULL AND JUNCTION BOXES SIZES SHALL COMPLY WITH N.E.O. ART. 314, VERIFY ALL SIZES WITH UNIT MANUFACTURER SHOP DRAWINGS AND FIELD REQUIREMENTS AND CONDITIONS BEFORE ROUGHING -IN. NOTES: 1. VERIFY ALL SIZES WITH UNIT MANUFACTURER SHOP DRAWING AND FIELD REQUIREMENTS AND CONDITIONS BEFORE ROUGHING -IN. REFER TO TO UNIT MANUFACTURER'S CONTROL DIAGRAMS. 2. ALL CONDUIT MAXIMUM LENGTHS SHALL COMPLY WITH MANUFACTURER REQUIREMENTS AS SHOWN ON UNIT MANUF DRAWNG% COORDINATE AT FIELD THE BEST ROUTING TO COMPLY. 3. RACEWAYS FILLING AND PULL AND JUNCTION BOXES SIZES SHALL COMPLY WTH N.E.C. ART. 314, C;1TfUIILOING Mt DEPT O ELE,CTHICAL SECTION A PPROVED 6y: L6 ' r,AT E SYMBOL V DEDICATED DIRECT TELEPHONE IN THE LOCATION OF THE VARIAN SY O OUTLET, REFER TO TELEPHONE AN Fes' VARIAN REQUIREMENTS REV 2 PG BEAM -ON WARNING LIGHT. PROVIDI O REQUIRED TO BLWK WHEN X -RAY U (24190, 250MA), INSIDE VAULT "I POINT IN THE CLINAC ROOM. PRO O WARNING LIGHTS RELAY A 1Y x 1. W CABINET. PROVIDE A POTTER -BRL O PROVIDE CCTV CAMERA SYSTEM 2 RECEPTACLES AS SHOWN. SYSTEM U.P.S. UNIT TO CAMERA AND TO C 0 PROVIDE TWO -WAY PATIENT MOAT O WITH PUSH -TO TALK CAPABILITIES. DOOR OPENER, 12OV, 1 OMOTORIZED g ELECTRICAL CONTRACTOR.. PROVIL 3�Q LASER POSITIONING LIGHT, POW-RE O CORD AND CONNECTED TO A OUR e RELAY J-BOX, PER UNIT MANUFAC LL. = M IN -ROOM MONITOR , PROVIDE A SI O VERIFY SIZE REQUIREMENTS AT FIE W FRIO PER DIRECTIONS MEDICAL EC O CONTROL EQUIPMENT PULL BOX 18 FROM PHYSICAL DAMAGE AND LOC COMBO TELEPHONE/DATA SIGNAL 10 VOICE AND DATA DWG E -4. SEE ALL SERVERS AND WORKSTATION E 11 COMBO TEIEPHONE/DATA SIGNAL 1 VOICE AND DATA DWG E -4. ON -BOARD IMAGER MAIN CIRCUIT 12 CONSOLE CONTRACTOR SHALL U INSTALLED IN A LOCKOUT /TAGOUT MANUfACTUREO BY G.E. CATJ OBI+ MAIN CIRCUIT BREAKER, IN NEMA ' f3 SHALL LABEL DISCONNECT AS 'MA 150A UNDER VOLTAGE RELEASE BF MANUFACTURER INSTALLATION GUN PROVIDE A 1Yx2Vx10' FLOOR RE, 14 RECOMMENDATIONS SEE DETAIL RE PROVIDE PROTECTION TO PREVENT EXTERNAL EMERGENCY SHUT -OFF 15 MANUAL RESET) IN ADDITION TO T WIRING CONTROL DIAGRAMS. ROOM SINGLE MASTER SWITCH WAIT ig TELEVISION CAN HAVE AN OPERAT CAN BE SUBORDINATED TO MIX OPERATION, REFER TO MANUFACTL 17 SAFETY DOOR INTERLOCK SWITCH. NORMALLY OPEN TYPE SWTCHES. t8 INDEPENDENT RADIATION DETECTOF 1.50 J -BOXES AT DETECTOR AND LOCATIONS BEFORE ROUGHING -IN NOTES: 1. CONTRACTOR SHALL PROVIDE ALL REQUIRED PROVISIONS FOR THE COMPLETE INSTALLATION OF THE WARNING LIGHTS SYSTEM, LASER POSITIONING LIGHTS AND RELAY, CCTV SYSTEM, INTERCOM, DETECTION SYSTEM, UNDER VOLTAGE RELAY, DOOR INTERLOCKS, EM SHUT -OFF SW, RM LTS RELAY, AND CLWAC PERIPHERALS AS SIHOWN. CONTRACTOR SHALL PULL VARIAN SUPPLIED CABLES WHEN APPLICABLE, REFER TO INTERCONNECTION WRING DIAGRAM FOR SILHOUETTE EDITION CLINAC REV 3, PG 7.38.0 AND SHOWN ON CUSTOMER SUPPLIED WE COORDINATE WITH MEDICAL EQUIPMENT INSTALLER AT FIELD BEFORE ROUGHING -IN. 2. ALL RACEWAYS FILLING AND PULL AND JUNCTION BOXES SIZES SHALL COMPLY WITH N.E.O. ART. 314, VERIFY ALL SIZES WITH UNIT MANUFACTURER SHOP DRAWINGS AND FIELD REQUIREMENTS AND CONDITIONS BEFORE ROUGHING -IN. NOTES: 1. VERIFY ALL SIZES WITH UNIT MANUFACTURER SHOP DRAWING AND FIELD REQUIREMENTS AND CONDITIONS BEFORE ROUGHING -IN. REFER TO TO UNIT MANUFACTURER'S CONTROL DIAGRAMS. 2. ALL CONDUIT MAXIMUM LENGTHS SHALL COMPLY WITH MANUFACTURER REQUIREMENTS AS SHOWN ON UNIT MANUF DRAWNG% COORDINATE AT FIELD THE BEST ROUTING TO COMPLY. 3. RACEWAYS FILLING AND PULL AND JUNCTION BOXES SIZES SHALL COMPLY WTH N.E.C. ART. 314, C;1TfUIILOING Mt DEPT O ELE,CTHICAL SECTION A PPROVED 6y: L6 ' r,AT E ul3 3� a U O` l c LL.. V W Fes' S U W (� N O W w > H 0 P4 U N N Z `/ U 3�Q w0co e LL. = M Z U C4 W v Z�~� W 0 CO VJ ul3 3� a U O` l c LL.. 0 VJ Z W (� N O W w > H 0 P4 `/ U w w w0co e LL. = M Z U Q Z�~� W Q CO VJ 00 0-9 0 12l DATE: 03/14/07 PROD NO: - DWG BY: JEB REV: - 04/06/07 A 212/10/2007 IT- REVISION 02/04/2008 3 CT -CHG. E -3 a