14-62-014I _ {
DEMOLITION NOTES: --
1O. REFER TO TOMOTHERAPY EQUIPMENT INSTALLATION DOCUMENTS PROVIDED <D
BY ACCURAY INCORPORATED FOR EXACT REQUIREMENTS AND LOCATION OF
DEVICES.
O EXISTING TOMOTHERAPY EQUIPMENT ELECTRICAL SERVICE TO BE MODIFIED RAD/ ONC Ell 0 11 1 RAID ONC.
AND REUSED AS INDICATED ON PLANS.
MARIA MED. DIR. ALLIE PHYS
O APOLLO LASER RECEPTACLE OUTLET TO REPLACE EXISTING. REUSE AMELIA STEVEN GARCIA ASSIST/
EXISTING DEDICATED CIRCUIT FROM CURRENT INSTALLATION. CIRCUIT 11-20 -7. RODRIGUES OLSZEWSKI STAFF SERRA WORK
SEE ACCURAY INC. DOCUMENTS. TOILETS
O EXISTING EMERGENCY OFF BUTTON SWITCH "EO" LOCATED IN CONTROL r
AREA STATION TO BE REMOVED AND REPLACED BY A NEW "EO" SWITCH AND
COVER PLATE TO MATCH AVAILABLE JUNCTION BOX. REUSE EXISTING
JUNCTION BOX.
—E�O5 PROVIDE A NEW 480V, 3 POLE, 70A SHUNT TRIP BREAKER IN NEMA I rl
ENCLOSURE TO REPLACE EXISTING 3P, 100A BREAKER LOCATED IN
TOMOTHERAPY EQUIPMENT CONTROL AREA TO SERVE NEW EQUIPMENT PDU.
O EXISTING 12" WIDE X 4" DEEP ELECTRICAL TROUGH WITH THREE SEPARATE
DIVIDED COMPARTMENTS FROM GANTRY TO JB1 TO BE REUSED BY NEW
INSTALLATION. POWER AND MISCELLANEOUS LOW VOLTAGE SYSTEMS `
CABLING SHALL RUN IN DEDICATED COMPARTMENTS PER NEC.
CORRIDOR
O POWER DISTRIBUTION UNIT (PDU), POWER CONDITIONER AND DATA SERVER
UNIT TO BE PROVIDED BY ACCURAY, SEE INSTALLATION DOCUMENTS FOR
RELEVANT INFORMATION.
O8 EXISTING DOOR SWITCH, RADIATION WARNING LIGHTS, LIGHT SWITCHES, '
JUNCTION BOXES, CONDUITS AND FLOOR TROUGHS ASSOCIATED WITH ' ;
CURRENT TOMOTHERAPY EQUIPMENT TO BE REMOVED SHALL BE REUSED AS r o
PART OF NEW UNIT INSTALLATION. PROVIDE NEW DEVICES, COVER PLATES ' '
SEE ENLARGED ,` le,�
AND WIRING AS NECESSARY. PLAN ON SHEET SOIL
O9 NEW POWER DISTRIBUTION UNIT "PDU" TO REPLACE EXISTING AT THIS SAME , E5.00 UTILITY MED
LOCATION. '
ROOM
10. EXISTING JUNCTION BOX #1 TO REMAIN (IN FRONT OF "PDU" TO REMAIN OPEN
FOR WIRING CONNECTIONS. o
11. EXISTING JUNCTION BOX JB2 LOCATED IN FLOOR BELOW PATIENT TABLE WITH EXAM EXAM EXAM EXAM
POWER / SIGNAL COMPARTMENT BARRIER. ' "- ' "- "' "'�'� DARK
12. EXISTING JUNCTION BOX JB3 LOCATED IN CEILING SPACE. CONTRACTOR TO ' MECHANICAL ROOM
VERIFY LOCATION AT FIELD AND REUSE. ' ELECT. STOR.
13. EXISTING DORADO LASER LOCATIONS TO REMAIN. a
1
14. EXISTING APOLLO LASER LOCATIONS TO REMAIN.
15. EXISTING CONVENIENCE TYPE RECEPTACLES IN TOMOTHERAPY AREA TO
Kz REMAIN. CONTRACTOR TO PROVIDE NEW DEVICES AND FACEPLATES TO
MATCH EXISTING. CIRCUITS TO BE ENGRAVED IN FACEPLATES.
16 EXISTING 4" PHYSICS CONDUIT JUNCTION BOX IN CONTROL ROOM AREA TO
REMAIN. CORRIDOR
17. EXISTING CCTV CAMERAS TO REMAIN.
14 15 15 ILE7, -I I
I
r 16
- - - - - -- -
GFI
_ LJ
O I1-20-9 i
10 '
EXISTING 4' PHYSICS ILE7; II
1 6 a;;; 1 CONDUIT TO REMAIN
i , 15
, !III , - - - - -- ,-
13 ------ - - - - -- .Iiiii ----- '----- � - LCJ I I
.I _ 1 ` 15
I'll ILE7 -9 t ey , (TYP.)
ILE7 -9 DS
1
13
(TYP.)
V�i��ABOVE ---- - - - - -- --- i- - - - - -- ABOVE ,)QiV
,�.J,' BELOW ; BELOW \Jill
ire
1 DEMOLITION FLOOR PLAN - ELECTRICAL
E3.00 SCALE: 1/4 " =1' -0"
iv; ,Tn IL20 -II
(TYP.)
ILE7 -9 (i
IL20 -11
�_
NUR
STATIC
r
- - - - r00NTFcOLL
SIMULATC)d
ARCHITECT:
BERMELLO AJAMIL & PARTNERS, INC.
Architecture Engineering Planning
- InteriorDesign Landscape Architecture
261_ Sz)uth Bayshore Drive 10th Floor
Miami, Florida 33133
(305) 859 -2050 Fax (305) 860 -3700
LOUIS J. AGUIRRE & ASSOCIATES, P.A.
Consulting Engineers
9150 South Dadeland Blvd. • Suite 900
Miami • Florida 33156 • EB0000924
Telephone 305.670.0141
Facsimile 305.670.0144
www.liaal2a.com
Project No.: 23075
Rw
• S �
L I$;cl•z2 '•QR.E.
L NH`' 116.42
CONSTRUCTION
DOCUMENTS
REVISIONS:
PROJECT NAME:
SMH -Tomo Equipment Replace
PROJECT ADDRESS:
7400 S.W. 62nd Ave.
South Miami, Florida 33143
KEY PLAN
COMM. NO. 40073.000
SCALE: AS SHOWN
DATE: 11-14-2013
DRAWN: AP
CHECKED: AG
CADD FILE: MINSTITUTIONAL
DRAWING TITLE:
DEMOLITION
FLOOR PLAN
ELECTRICAL
SHEET NO.
E3.00