DEVICES TO REMAIN
TO NEW WORK PLAN
DEVICES.
ENLARGED CONTROL AREA - DEMOLITION - ELECTRICAL_
i
1L16 -36 12
iii *0'3"
ENLARGED CONTROL AREA - NEW- WORK------ ELECTRICAJ
111111111 1 M I -I
1
KEYED NOTES:
0 48OV, 3- PHASE, 60A OBI MAIN CIRCUIT BREAKER DISCONNECT. PROVIDE GE #OB16OA48OV.
ENSURE 42' CLEARANCE IS MAINTAINED IN FRONT OF PANEL PER NEC ART 110.26.
0 INTER CONNECT PANEL - PROVIDED BY VARIAN.
0 HF CABINET BY VARIAN
0 3.51X10' SURFACE ELECTRICAL DUCT WITH TWO PARTITIONS FROM IDP ON WALL TO CLINAC
STAND ON FLOOR. ELECTRICAL DUCT SHALL BE WALKER WALLDUCT FOR HEALTHCARE.
(WD10W350 SERIES).
0 PROVIDE 2 -4' CONDUIT FOR LOW VOLTAGE. TERMINATE ONE END TO WIRING ACCESS BEHIND CONTROL
COUNTER AND THE OTHER END INTO THE SURFACE ELECTRICAL DUCT TO THE REAR
OF THE CLINAC STAND.
0 EXISTING 1' SPARE CONDUIT TO BE USED FOR NEW 480V OBI FEEDER.
0 EXISTING SWITCHES AND DOOR PUSH BUTTON TO BE RELOCATED TO ACCOMODATE WORK
TO WALL - REFER TO NEW WORK PLAN FOR NEW LOCATIONS.
0 NEW 'OBI IN USE' WARNING LIGHT. HEALTHCARE LIGHTING SYSTEMS #HSG301- XIU -R. PROVIDE
RELAY AND CONNECT TO OBIPER VARIAN VENDOR DRAWINGS. CONNECT TO ADJACENT
WARNING LIGHT'S 120V SOURCE.
0 CONNECT TO EXISTING SHUNT TRIP PUSHBUTTONS IN ROOM WITH 2 #12 IN1 /2C.
�o REPLACE EXISTING RECEPTACLES WITH GFI TYPE.
>i REMOVE EXISTING RECEPTACLE. REUSE EXISTING CIRCUIT FOR NEW DUPLEX RECEPTACLES
IN NEW CASEWORK. REFER TO NEW WORK PLAN ON THIS SHEET.
t2 RECONNECT TO EXISTING CIRCUIT PREVIOUSLY FEEDING DEMOLISHED RECEPTACLE.
I
I
ti
4
3
SHOWN FOR REFERENCE
ONLY - TYPICAL
PARTIAL FLOOR PLAN - NEW WORK - ELECTRICAL_
111111111 1 a I
4
5
5
CONSTRUCTION LEGEND
DESCRIPTION
rN,EW
EXISTING TO BE REMOVED
EXISTING TO REMAIN
NEW CONSTRUCTION
CONNECT TO EXIST. AT THIS POINT
BERMELLO AJAMIL & PARTNERS, INC.
Architecture Engineering Planning
Interior Design Landscape Architecture
2601 South Bayshore Drive 10th Floor
Miami, Florida 33133
(305) 859 -2050 Fax (305) 860 -3700
M
SOUTH MIAMI
HOSPITAL
6200 SW 73 STREET, MIAMI, FLORIDA. 33143
- r
Smith
Ss?t Seckman
Reid, Inc.
600 West Hillsboro, Suite 300
Deerfield Beach, FL 33441
(954) 421 -1260
FAX: (954) 421 -1466
vvww.ssr- inc.com
Firm Certification - E13- 0000857
Alfred Esquivel PE - 45863
Bruce W. Peek PE - 35901
Donald Netinho PE - 48454
Erik S. Walter PE -63929
John E. Moulder PE - 33200
Michael A. Steward PE - 65701
Usman A.Tariq PE - 66485
SSR JOB # 0732091,0
°a
».
s
v ,
0NARKS N&I
PROJECT
SOUTH MLAA4I HOSPITAL
LINEAR ACCELERATOR
OBI UPGRADE
..O ADD
6200 SW 73 STREET
Miami, Florida ..•.
COMM. NO.
SCALE: AS SHOWN
DATE: 12 -20 -2007
DRAWN:
CHECKED:
A CADD FILE:
DRAWING TITLE:
PARTIAL FLOOR
PLAN - NEW WORK
ELECTRICAL
SHEET NO.
E2.1