13-1218-041
PROPOSED DENTAL ELECTRICAL SPECIFICATIONS ann - ro. ELECTRICAL NOTES 1. THIS SPECIFICATION SHEET IS INTENDED AS A GUIDE FOR TRADESMEN. THE FLOOR PLANS ENCLOSED HEREIN ARE SUGGESTIONS FOR THE PLACEMENT OF DENTAL EQUIPMENT. THEY ARE NOT INTENDED FOR CONSTRUCTION, 2. EXACT EQUIPMENT LOCATIONS MUST BE JOB SITE VERIFIED BY THE HENRY SCHEIN DENTAL EQUIPMENT SPECIALIST, 3. FOLLOW MANUFACTURER'S DRAWINGS FOR EXACT REQUIREMENTS FOR ANY EQUIPMENT SUPPLIED BY HENRY SCHEIN DENTAL. CONSULT WITH HENRY SCHEIN DENTAL REP FOR ADDITIONAL INFORMATION. 4. GFCI PROTECTION OR REDUNDANT GROUND IN DENTAL CHAIR RECEPTACLES, DENTAL UTILITY CABINETS AND ANY OTHER AREAS REQUIRED BY LOCAL CODE IS THE RESPONSIBILITY OF THE ELECTRICAL CONTRACTOR. 5. ADDITIONAL CONVENIENCE OUTLETS REQUIRED ON JOBSITE ARE THE RESPONSIBILITY OF ELECTRICAL CONTRACTORS. FOLLOW LOCAL CODE RESTRICTIONS. 6. ALL COMPUTER NETWORKING AND WORKSTATIONS MUST BE SPECIFIED BY CUSTOMER'S COMPUTER SUPPLIER. IF HENRY SCHEIN WILL BE SUPPLYING THE COMPUTER NETWORKING SYSTEM, SEE SHEET SE.2 FOR THE OFFICE AUTOMATION PLAN. ELECTRICAL LEGEND NEW DUPLEX OUTLET. HEIGHT TO BE SPECIFIED ON -SITE UNLESS NOTED OTHERWISE, NEW ETHERNET JACK LOCATION. HEIGHT TO BE SPECIFIED ON -SITE UNLESS NOTED OTHERWISE. SEE SHEETS SPE.1 & SPE.2 FOR ILLUSTRATIONS & DETAILS ELECTRICAL SPECIFICATIONS _ PLEASE NOTE: ALL REQUIREMENTS TO BE VERIFIED BY MANUFACTURER'S SPEC UTILITIES SHEETS PROVIDED BY HENRY SCHEIN DENTAL EQUIPMENT SPECIALIST. INT.SQ.FT.= BY OTHERS MANUFACTURER'S SPECS SUPERSEDE ANY AND ALL INFORMATION CONTAINED r HEREIN. o N E5 O J < o N w yw ( u O < �w UC'1 W 0 uj t�Ow J N z � z F- U >O = N DESCRIPTION uj 0 0 0 0 Lu uj 0 5 1 UTILITY CENTER - CONTRACTOR TO PROVIDE REQ'D UTILITES. BACK FLOW PREVENTION • 115V, 20A 1 REQUIRED AS PER LOCAL CODE. 22 3 UTILITY CENTER - CONTRACTOR TO PROVIDE REQ'D UTILITES. RACK FLOW PREVENTION + 115V, 20A 3 REQUIRED AS PER LOCAL CODE. 20 4A 11 1/4" CONDUIT- PIPE CHASE WITH PULL - STRING SUPPLIED AND INSTALLED BY CONTRACTOR. 17 5 DENTAL CHAIR RECEPTACLE - SUPPLIED & INSTALLED BY CONTRACTOR, EXACT LOCATION TO 115V, 20A 5 BE SPOTTED BY HENRY SCHEIN DENTAL EQUIPMENT SPECIALIST. 2 6 DENTAL TRACK LIGHT - CONTRACTOR TO PROVIDE REQ'D UTILITIES AND MOUNTING SUPPORT. • 115V, 15A 6 LIGHT SUPPLIED AND INSTALLED BY HENRY SCHEIN DENTAL. 20 6A DENTAL CEILING LIGHT - CONTRACTOR TO PROVIDE REQ'D UTILITES AND MOUNTING SUPPORT. • 115V, 15A 6A LIGHT SUPPLIED AND INSTALLED BY HENRY SCHEIN DENTAL. WALL MOUNTED INTRAORAL X -RAY HEAD, ARM, AND WALL BRACKET - CONTRACTOR TO 5 9 PROVIDE REQ'D UTILITES AND MOUNTING SUPPORT. X -RAY UNIT SUPPLIED AND INSTALLED BY • 115V, 15A • ? HENRY SCHEIN DENTAL. REQUIRES A 3 WIRE GROUNDED CIRCUIT. X -RAY MASTER CONTROL - CONTRACTOR TO SUPPLY AND INSTALL PIPE CHASE WITH 5 ?A PULL - STRING TO (9) LOCATION WHEN (9A) IS REMOTE. VERIFY SUPPLIER OF WIRING WITH HENRY • SCHEIN EQUIPMENT SPECIALIST. FINAL CONNECTION BY HENRY SCHEIN DENTAL. X -RAY REMOTE EXPOSURE SUTTON - CONTRACTOR TO SUPPLY AND INSTALL PIPE CHASE WITH 5 9B PULL - STRING TO (9A) OR (9C) LOCATION. VERIFY SUPPLIER OF EXPOSURE SUTTON AND WIRING • 99 WITH HENRY SCHEIN EQUIPMENT SPECIALIST. FINAL CONNECTION BY HENRY SCHEIN DENTAL. PANORAMIC AND /OR CEPHALOMETRIC XRAY/ CONE SEAM UNIT - CONTRACTOR TO PROVIDE 115V & 230V, 1 9C REQ'D UTILITES. REQUIRES A 3 WIRE GROUNDED CIRCUIT. X -RAY UNIT SUPPLIED AND INSTALLED • 20A • ?C BY HENRY SCHEIN DENTAL. 4 14A STERILIZER (AUTO- CLAVE) - CONTRACTOR TO PROVIDE REQ'D UTILITES. STERILIZER SUPPLIED AND • 115V, 15A • 14A INSTALLED BY HENRY SCHEIN DENTAL. VERIFY IF SWITCHED OUTLET IS REQ'D. 2 15 ULTRASONIC CLEANER - CONTRACTOR TO PROVIDE REQ'D UTILITIES. CLEANER SUPPLIED AND + 115V, 15A 15 INSTALLED BY HENRY SCHEIN DENTAL. 2 16 MODEL TRIMMER - CONTRACTOR TO PROVIDE REQ'D UTILITIES. MODEL TRIMMER SUPPLIED AND + 115V, 15A 16 INSTALLED BY HENRY SCHEIN DENTAL. BACK FLOW PREVENTION REQUIRED PER CODE. EXHAUST FAN FOR DENTAL MECHANICAL CLOSET - SUPPLIED AND INSTALLED BY CONTRACTOR. 1 23A A THERMOSTAT CONTROLLED LOW SONE FAN IS REQUIRED TO KEEP ROOM TEMPERATURE • 115V, 20A s WITHIN THE EQUIPMENT MANUFACTURERS RECOMMENDED OPERATING TEMPERATURE RANGE. VENT TO OUTSIDE. DENTAL AIR COMPRESSOR (AIRSTAR 50) - FOR GAS -DRIVEN DEVICES IN A LEVEL 3 FACILITY CONTRACTOR TO PROVIDE REQ'D UTILITES AND FRESH -AIR INTAKE VENT. 1/2" I.D. COPPER TYPE 1 25 K OR L SUPPLY LINES TO LOCATIONS THAT REQUIRE AIR. SUCK BOOST TRANSFORMER SUPPLIED • 230V, 30A • • • 25 AND INSTALLED BY CONTRACTOR (IF REQ'D). FINAL CONNECTIONS BY CONTRACTOR. REFER TO NFPA 99, 2005, 5.3.3.5 FOR REGULATIONS & REQUIREMENTS. DENTAL AIR COMPRESSOR (AIRSTAR 70) - FOR GAS- DRIVEN DEVICES IN A LEVEL 3 FACILITY. CONTRACTOR TO PROVIDE REQ'D UTILITES AND FRESH -AIR INTAKE VENT. 1/2" I.D. COPPER TYPE 1 25 K OR L SUPPLY LINES TO LOCATIONS THAT REQUIRE AIR. BUCK BOOST TRANSFORMER SUPPLIED • 230V, 40A • • • 25 AND INSTALLED BY CONTRACTOR (IF REQ'D). FINAL CONNECTIONS BY CONTRACTOR. REFER TO NFPA 99, 2005, 5.3.3.5 FOR REGULATIONS & REQUIREMENTS. VACUUM PUMP (DRY) - CONTRACTOR TO PROVIDE REQ'D UTILITIES. SCH 40 PVC MAIN TRUNK LINES (WHERE PERMITTED BY CODE). TRUNK, REDUCTION AND BRANCH LINE SIZES PER MFR RECOMMENDATIONS. EXHAUST PUMP TO OUTSIDE. SACK FLOW PREVENTION REQ'D PER 2 27A LOCAL CODE. COLLECTION CANISTER (CAS) SUPPLIED BY HENRY SCHEIN DENTAL, INSTALLED • 230V, 20A • • • 27A BY CONTRACTOR. BUCK BOOST TRANSFORMER SUPPLIED AND INSTALLED BY CONTRACTOR (IF REQ'D). FINAL CONNECTIONS BY CONTRACTOR. PROVIDE WATER LINE WITH HOSE SIB NEAR UNIT. WATER SOLENOID W /FILTER - CONTRACTOR TO PROVIDE REQ'D UTILITIES. SOLENOID SUPPLIED 1 30A BY HENRY SCHEIN DENTAL, INSTALLED BY CONTRACTOR ON COLD WATER SUPPLY LINE (FOR • 115V, 15A • • 30A DENTAL EQUIPMENT ONLY). REMOTE PANEL (2 AIR, WATER, & 2 VAC) - SUPPLIED BY HENRY SCHEIN DENTAL, INSTALLED BY 1 32A CONTRACTOR 60" A.F.F. CONTRACTOR TO SUPPLY AND INSTALL LOW VOLTAGE WIRING PER • 32A MFR SPECS. N20 -02 SYSTEM MANIFOLD - CONTRACTOR TO PROVIDE REQ'D UTILITIES. MANIFOLD SUPPLIED 1 33 BY HENRY SCHEIN DENTAL, INSTALLED BY CONTRACTOR. SILVER SOLDERED K OR L COPPER + 115V, 20A 33 TRUNK LINES RUNNING TO EACH TREATMENT ROOM. CONTRACTOR TO PRESSURE TEST LINES SEPARATELY WITH DRY NITROGEN AT 150 PSI FOR 24 HOURS. FOLLOW NFPA CODES. 1 34 N20 -02 SYSTEM ALARM - CONTRACTOR TO PROVIDE REQ'D UTILITES. ALARM SUPPLIED BY + 24V • 34 HENRY SCHEIN DENTAL, INSTALLED BY CONTRACTOR. 1 46 CUBEX STORAGE UNITS - CONTRACTOR TO PROVIDE REQ'D UTILITIES. UNITS SUPPLIED AND + 11OV,20A + 46 INSTALLED BY HENRY SCHEIN DENTAL. > u V J w. Z w Z t� J N O (y) Z ..J •[ c c_/3 op LLJ < co z O F- WN c/) o 0- w Q O 3: pp a z w X �p U w U) O C� (y) u.a (y) M-- d V w C p¢ Z:�i V) O N V) cr I-0 Z O U 0 HENRY SCHEIN REP: JOHN TURNER CENTER: _WEST PALM BEACH PHONE #: 800 645 -6594 w Oz0C?Lu< O w -K w cn u- 0 X400 �w U_ U w F n w w V) 0 0 < ? w ua zCO�� < LL ( Q�ZOoz O zw3u LLjLU 0whz¢w�" Lz 0 0 0 Wz p¢0 < °w0�n � � w�Vw < 0 0 z < :D 0 < �5 Zw� cn Za�('l u co Z z ) 2V o O zZ in U z w s> O z 00 °U¢ w ©�NCn�Z w w � w _ z_ ui �a t 0 © W .J w DRAWING NAME: SOUTHMIAMI -F PROJECT START DATE: 07/02/2013 _ FINALS START DATE: 09/10/2013 DRAWN BY: OTHERS FINALS BY: TAS CHECKED BY: JBK REVISIONS: _ INT.SQ.FT.= BY OTHERS SCALE: SHT, SIZE, 3/1611= 1 ' -0" D PLAN 7 Z LL cw " } z LL S c Cr C 4